Keysight B1500A Configuration and Connection Guide, Edition 6
3-13
Connection Guide for Wafer Prober
Connection for low current measurement
Figure 3-13
Guarding
WARNING
Do not touch the guard terminal with bare hands because you may be shocked by
high voltage. The potential of the guard terminal is equal to the output voltage.
CAUTION
Never connect the Guard terminal to any other output, including circuit common, chassis
ground, or any other guard terminal. Doing so will damage the B1500A.
A
Buffer
Guard
Force
DUT