Frequency Sweep Commands
4
TRIGger:SOURce
{
IMMediate|EXTernal|BUS
}
TRIGger:SOURce?
Select the source from which the function generator will accept a trigger.
The function generator will accept an immediate internal trigger, a
hardware trigger from the rear-panel Trig In connector, or a software
(bus) trigger. The default is IMM. The
:SOUR?
query returns “IMM”,
“EXT”, or “BUS”
• When the Immediate (internal) source is selected, the function
generator outputs a continuous sweep at a rate determined by the
sweep time specified (
SWE:TIME
command)
plus
1 ms.
• When the External source is selected, the function generator will
accept a hardware trigger applied to the rear-panel Trig In connector.
The function generator initiates one sweep each time Trig In receives
a TTL pulse with the edge polarity specified by the
TRIG:SLOP
command (see page 209). Note that the trigger period must be greater
than or equal to the specified sweep time
plus
1 ms.
• When the Bus (software) source is selected, the function generator
initiates one sweep each time a bus trigger command is received.
To trigger the function generator from the remote interface (GPIB,
USB, or LAN), send the
*TRG
(trigger) command. The front-
panel
key is illuminated when the function generator is waiting
for a bus trigger.
• The
APPLy
command automatically sets the trigger source to
Immediate (equivalent to
TRIG:SOUR IMM
command).
• To ensure synchronization when the Bus source is selected, send the
*WAI
(wait) command. When the
*WAI
command is executed, the
function generator waits for all pending operations to complete before
executing any additional commands. For example, the following
command string guarantees that the first trigger is accepted and the
operation is executed before the second trigger is recognized.
TRIG:SOUR BUS;*TRG;*WAI;*TRG;*WAI
• You can use the
*OPC?
(operation complete query) command or the
*OPC
(operation complete) command to signal when the sweep is
complete. The
*OPC?
command returns “1” to the output buffer when
the sweep is complete. The
*OPC
command sets the “Operation
Complete” bit (bit 0) in the Standard Event register when the sweep
is complete.
33210A users guide.book Page 208 Wednesday, July 16, 2008 11:16 AM
Содержание 33210A
Страница 1: ...Keysight 33210A 10 MHz Function Arbitrary Waveform Generator User s Guide ...
Страница 2: ......
Страница 3: ......
Страница 14: ...12 Contents Contents 33210A users guide book Page 12 Wednesday July 16 2008 11 16 AM ...
Страница 15: ...1 1 Quick Start 33210A users guide book Page 13 Wednesday July 16 2008 11 16 AM ...
Страница 33: ...2 2 Front Panel Menu Operation 33210A users guide book Page 31 Wednesday July 16 2008 11 16 AM ...
Страница 53: ...3 3 Features and Functions 33210A users guide book Page 51 Wednesday July 16 2008 11 16 AM ...
Страница 147: ...4 4 Remote Interface Reference 33210A users guide book Page 145 Wednesday July 16 2008 11 16 AM ...
Страница 273: ...5 5 Error Messages 33210A users guide book Page 271 Wednesday July 16 2008 11 16 AM ...
Страница 301: ...6 6 Application Programs 33210A users guide book Page 299 Wednesday July 16 2008 11 16 AM ...
Страница 311: ...7 7 Tutorial 33210A users guide book Page 309 Wednesday July 16 2008 11 16 AM ...
Страница 335: ...8 8 Specifications 33210A users guide book Page 333 Wednesday July 16 2008 11 16 AM ...