Installation on 16800 Series and 16901A Logic Analysis Systems
2
Step 1. Prepare the instrument for disassembly
Do this procedure before doing any disassembly procedure on the
instrument.
1 Close the
Keysight Logic Analyzer
application software.
2 Back up any data and license files you want to retain using LAN, USB
flash drive, external USB CD/DVD writer, etc.
3 Power off the system using one of the methods explained below.
4 Remove the power cord.
5 Move the instrument to a static safe work environment.
To power off the system
There are several ways to power off the logic analyzer.
Using the Windows Shutdown
On the logic analyzer desktop click
Start> Shut Down
or if you are running
remote desktop, click
Start> Settings> Windows Security> Shut Down
. This
software power off does the following:
• Closes all programs that are running.
WARNING
Hazardous voltages exist on the power supply. To avoid electrical shock,
disconnect the power from the instrument before performing the
following procedures. After disconnecting the power, wait at least six
minutes for the capacitors on the power supply board to discharge
before servicing the instrument.
CAUTION
Damage can occur to electronic components if you remove or replace
assemblies when the instrument is on or when the power cable is
connected. Never attempt to remove or install any assembly with the
instrument on or with the power cable connected.
NOTE
When powering off the logic analyzer, wait until the fans stop turning
(about 15 seconds) before turning the logic analyzer back on. This
ensures that internal circuitry restarts in a known state.