Z5623A Option H01
General Specifications
General Specifications
Specifications for the Z5623A Option H01 Dynamic Accuracy Test Set is characteristic for the
System performance of the PNA and Test Set. Actual performance of the system is based on
the customers PNA that is used with the
T
est
S
et.
A functional certificate is supplied for the Z5623A Option H01.
The Z5623A Option H01
performance is based on external components such as the calibration kit and network analyzer.
There are no internal adjustment in the Z5623A Option H01, therefore an annual calibration is
not required.
This product has an autoranging line voltage input. Be sure the supply voltage is
within the specified range. If the AC line voltage does not fall within these
ranges, an autotransformer that provides third wire continuity to earth ground
should be used.
Power Requirements
Verify that the required ac power is available at all necessary locations before installing the Test
Set to the PNA.
• 100/120/220/240 V (50/60 Hz)
• Air conditioning equipment (or other motor–operated equipment) should not be placed on
the same ac line that powers the Test Set.
• The instruments can operate with mains supply voltage fluctuations up to ± 10% of the
nominal voltage.
• Z5623AH01 maximum power is 40 Watts
This is a Safety Class I Product (provided with a protective earthing ground incorporated
in the power cord). The mains plug shall be only be inserted in a socket outlet provided
with a protective earth contact. Any interruption of the protective conductor inside or
outside of the product is likely to make the product dangerous. Intentional interruption is
prohibited