Z5623A Option H01
Front and Rear Panel Features
Rear Panel
Figure
5
Rear Panel
GPIB Connector
This connector allows the test set to be connected directly to a controller.
Address Switch
The address switch sets the GPIB address of the test set. See
Line Module
The line module contains the power cable receptacle and the line fuse. The line module is an
autoranging input and is designed to be used with an ac power source with a nominal voltage of
either 115 V or 230 V.
Power Cables
The line power cable is supplied in one of several configurations, depending on the destination
of the original shipment.
Each instrument is equipped with a three-wire power cable. When connected to an appropriate
ac power receptacle, this cable grounds the instrument chassis. The type of power cable
shipped with each instrument is dependent on the country of destination. See
for the part numbers of these power cables. Cables are available in different lengths.
“Keysight Support, Services, and Assistance” on page 33
numbers of cables other than those described in
Always use the three-prong AC power cord supplied with this product. Failure to
ensure adequate grounding by not using this cord may cause damage to the
product.
This is a Safety Class I product (provided with a protective earthing ground
incorporated in the power cord). The mains plug shall only be inserted into a
socket outlet provided with a protective earth contact. Any interruption of the
protective conductor, inside or outside the instrument, is likely to make the
instrument dangerous. Intentional interruption of the protective cover is
prohibited.
GPIB Connector Address Switch
Line Module
Parallel In/Out Ports
(Non-Functional)
10
Keysight Z5623A-H01 User's Guide