Z5623A Option H01
Electrical and Environmental Requirements
Required Conditions for Accuracy Enhanced Measurement
Accuracy–enhanced (error–corrected) measurements require the ambient temperature of the
PNA and Test Set to be maintained within ±1 °C of the ambient temperature at calibration.
Table 3
Performance Characteristics
Parameter
Typical
Frequency Range
1.2 GHz
Insertion Loss
1
1. Attenuator A = 0 dB, Attenuator B = 0 dB, Amplifier OUT
Source to Power Meter Port
–7.5 dB (± 3dB)
Source to Power Meter Port with
Amplifier
+9.0 dB (± 4dB)
Source to Receiver Port
–7 dB (± 3 dB)
Power Meter Port to Source
(Amplifier IN selected)
≤
–45 dB
Input Power Damage Level
>1 watt (RF + DC)
8
Keysight Z5623A-H01 User's Guide