384 Keysight CXG, EXG, and MXG X-Series Signal Generators Programming Guide
Creating and Downloading User–Data Files
Data Transfer Troubleshooting
Data Transfer Troubleshooting
This section is divided by the following data transfer methods:
“User File Download Problems” on page 384
“PRAM Download Problems” on page 385
“User FIR Filter Coefficient File Download Problems” on page 387
Each section contains the following troubleshooting information:
— a list of symptoms and possible causes of typical problems encountered
while downloading data to the signal generator
— reminders regarding special considerations and file requirements
— tips on creating data, transferring data, data application and memory usage
User File Download Problems
Data Requirements
— The user file selected must entirely fill the data field of each timeslot.
— The user file must be a multiple of 8 bits, so that it can be represented in
ASCII characters.
— Available volatile memory must be large enough to support both the data
field bits and the framing bits.
This section applies to vector models with an installed baseband generator
option.
Table 6-14
Use–File Download Trouble – Symptoms and Causes
Symptom
Possible Cause
At the RF output,
some data modulated,
some data missing
Data does not completely fill an integer number of timeslots.
If a user file fills the data fields of more than one timeslot in a continuously repeating framed
transmission, the user file will be restarted after the last timeslot containing completely filled data
fields. For example, if the user file contains enough data to fill the data fields of 3.5 timeslots,
firmware will load 3 timeslots with data and restart the user file after the third timeslot. The last
0.5 timeslot worth of data will never be modulated.