26
U9361C/F/G/M RCal Receiver Calibrator Measurement Guide
Demonstrations
5.
Select
Input/Output
>
Calibration
>
Configuration
to access the RCal Calibration
Configuration window.
6.
Select
Use Current Meas
, then select the
Calibrate
check box >
Calibrate Checked Rows
to
populate a row using the current measurement settings, and run the calibration.
7.
When calibration is complete, select
Apply
, then
Close
the window.
8.
Disconnect the cable from the RCal RF Out port and connect it to the signal generator RF Out
port.