U9361C/F/G/M RCal Receiver Calibrator Measurement Guide
39
Demonstrations
10.
Populate a row with the following parameters: (which should also be set in the signal analyzer
meas bar)
Type:
Complex
Start/Stop Freq:
5 GHz
Freq Points:
1
Mech Atten:
Step
Mech Atten Start:
10 dB
Mech Atten Stop:
10 dB
Elec Atten:
Bypass
IF Gain:
Auto
IF Path:
25 MHz
Preamp:
Off
Coupling:
AC
Phase Noise Opt:
Best Wide Offset
uW Path Ctrl:
uW Presel Bypass
11.
Run the calibration after selecting the calibrate check box.