16
U9361C/F/G/M RCal Receiver Calibrator Measurement Guide
RCal Software – Calibration and Application
The following message will be displayed during the calibration process:
When calibration is complete, you should see the following message which will disappear
automatically:
To cancel a calibration in progress, press the
Cancel
button or use the
Esc
key.
After calibration, you should see a time stamp for all calibrated rows under the
Last Calibrated
column.
Applying a Calibration
To apply the correction, select the
Apply
column in the row. The instrument will compare its current
state to all corrections in the active RCal correction pool and apply correction factors according to
the
Match State
rules. The
Applied
column should now show
Yes
, indicating that the selected
calibration is being applied.
When calibrations are active, the Corr CCorr (Complex Correction) option on the measurement
panel should show "RCal" in orange as shown below.
The Apply indicator is dynamic; as you change analyzer settings, RCal will re-evaluate to check if
any RCal corrections in the active correction pool are applicable.