Keysight U3042AM20 User’s and Service Guide
63
U3042AM20
Cal Kit Operational Check
Figure 44
Calibration Complete
7. Repeat
See “Connect the ECal or the mechanical cal kit to Port 1 or the
port you are testing. Torque to 8 in-lb. For further information, refer to the
See “At the Calibration Completed prompt,
select Save As User Calset and type the name 999.1. Overwrite the Calset if
it already exists and Save.” on page 62.
(1-Port Calibration Procedure) for
Ports 2 thru 24. When finished, there should be 24 Cal Sets saved with the
titles “999.1” thru “999.24” (24-Port).
If you are using an ECal module you can verify the individual port
calibration by selecting Response > Cal > More > ECal > ECal Confidence
Check. Click Change Measurement and select the test port S-Parameter >
Apply > OK > Read Module Data. For further information refer to the Help
menu.
8. Select Trace/Chan > Trace > Delete Trace. There should be no traces on
the analyzer’s display. If traces are listed, select each individually and
delete > OK.
9. To launch the Cal Set Viewer toolbar. Select Response > Cal > Manage Cals
> Cal Set Viewer.