Cal Kit Operational Check
-
1-Port Calibration and Verification Procedure
1. Connect the ECal or the mechanical cal kit to Port 1 or the port you are testing.
Torque to 8 in-lb. For further information refer to the Help menu.
2. Perform a 1-Port Calibration on Port 1. On the analyzer, select
Response > Cal > Start Cal > Calibration Wizard -OR-
(Response > Cal > Other Cals)
.
a. If using a mechanical cal kit, select SmartCal > (Guided Calibration)
>
Next.
b. If using an ECal module, connect the ECal to a analyzer’s USB port.
Select Use Electronic Calibration (ECal) > Next.
3. Continue following the Cal Wizard prompts. In the “Select Calibration Ports
and ECal Module” dialog box press Clear All, then select Port > Next
>
Measure.
4. Ensure the Cal Kit you are using is indicated on the right side of the dialog box.
Figure 43
1-Port Calibration
5. Continue to follow the prompts until the “Calibration Completed” dialog
box appears.
6. At the Calibration Completed prompt, select Save As User Calset and type
the name 999.1. Overwrite the Calset if it already exists and Save.
If you do not have a keyboard, select Save As User Calset > Edit Name and
save as 999.x. X is the port number you are calibrating. Use the numeric
keypad on the PNA-X front panel to enter “999.1.”
Confirm