60
Keysight U3042AM20 User’s and Service Guide
U3042AM20
Cal Kit Operational Check
-
Equipment Required
The U3042AM20 requires that the user be familiar with the equipment and
components listed.
Operational Check Procedure
The sequence of this procedure is very important and must be followed or the
performance accuracy and results may vary from the reference plots provided.
Preparing the N5242A
1. Connect the Test Set to the 4-Port PNA-X using the interconnect cables as
2. Turn On the Test Set.
3. On the PNA-X, select Response > Cal > Manage Cals > Cal Set. Delete or
Rename any Cal Sets titled “999.1” thru “999.24” (24-Port), although it is
unlikely that you will find Cal Sets with these names.
4. Verify that the PNA-X is in 24-Port. See the bottom of the measurement
window.
a. If only four S-Parameters are listed, select
Utility > System > Configure > Multiport Capability -OR-
(Instrument > Setup External Hardware > Multiport > Multiport
Configuration...)
. Select Restart as multiport PNA with this test set
and select U3042AM20 (24-Port) from the drop-down menu > OK.
Refer to
.
5. Press Preset.
6. Verify that the [Start Frequency] is set to [10 MHz].
7. Verify that the [Stop Frequency] is set to [26.5 MHz].
8. Verify that the [Power] is to set to [−5 dBm]. If not, select [Power] > Power
Level and enter [−5 dBm].
9. Select Response > Avg > IF Bandwidth > 100 Hz > OK.
Table 17
Equipment List
Description
Qty
N4691A 3.5 mm ECal Module 10 MHz - 26.5 GHz (Option 00F or M0F) or
N4691B 3.5 mm ECal Module 300 kHz - 26.5 GHz (Option 00F or M0F) or
Mechanical cal kit 85052B or 85052D
1
N5242A Option 400 and 551
1
Set of interconnect cables (PNA-X and Test Set), see
Interface Cable Connections (U3021-60077)” on page 22
.
1