N5222A and N5242A RF Interface Cable Connections (U3021-60077)
-
N5222A and N5242A RF Interface Cable Connections
(U3021-60077)
illustrates the cable configuration of the Test Set to the
analyzer. The cables have been supplied with the kit (U3021-60077).
Cable Connection Procedure
1.
Remove the SOURCE OUT to CPLR THRU and RCVR IN to CPLR ARM
jumpers (x8) on the analyzer. The RCVR R1 - R4 to SOURCE OUT reference
loop jumpers (x4) remain on the front panel.
2.
Connect the RF interconnect cables from the PNA or PNA-X to the test set
in the order listed. As you are connecting each cable, torque to 8 in-lb. The
longer, straight end of each cable is connected to the test set. Refer to
and
Over torque will cause damage to the test set and may cause connectors to
spin or become loose.
Table 7
PNA or PNA-X RF Interface Cable Connection (U3021-60077)
Numeric Order
RF Cables
From PNA-X
To Test Set
1
U3042-20122
Port 1 CPLR THRU
Port 5 CPLR THRU
2
U3042-20120
Port 1 SOURCE OUT
Port 5 SOURCE IN
3
U3042-20119
Port 1 CPLR ARM
Port 5 CPLR ARM
4
U3042-20121
Port 1 RCVR A IN
Port 5 RCVR OUT
5
U3042-20123
Port 3 CPLR ARM
Port 6 CPLR ARM
6
U3042-20123
Port 3 RCVR C IN
Port 6 RCVR OUT
7
U3042-20124
Port 3 SOURCE OUT
Port 6 SOURCE IN
8
U3042-20124
Port 3 CPLR THRU
Port 6 CPLR THRU
9
U3042-20123
Port 4 RCVR D IN
Port 7 RCVR OUT
10
U3042-20123
Port 4 CPLR ARM
Port 7 CPLR ARM
11
U3042-20124
Port 4 SOURCE OUT
Port 7 SOURCE IN
12
U3042-20124
Port 4 CPLR THRU
Port 7 CPLR THRU
13
U3042-20126
Port 2 CPLR THRU
Port 8 CPLR THRU
14
U3042-20128
Port 2 SOURCE OUT
Port 8 SOURCE IN
15
U3042-20125
Port 2 RCVR B IN
Port 8 RCVR OUT
Continued on page 23