U3022AH10
Address and Data Values
If the ports have different addresses, two separate address commands are needed.
P
ort 3 is the Source
(with Noise Source Path)
and Port
4
is the Receiver.
Source Port 3 = address 0, data
9
, and Receiver Port
4
= address
16
, data
16
.
Two separate commands must be sent. You may use the same dialog box. The entry
will be 0.
9
in one command line, and
16.16
in the second line.
Example 2
, 2b
:
(
STD or
Address and Data Values
Setting the Test Port Paths with Address and Data
for
information to set the internal switch paths of the Test Set.
The address is the first value in the Test Set I/O control or GPIB data command. The second
value controls the source and receiver paths of the ports. To select a Test Set port
configuration, both switches must be set to complete the source and receiver paths. To do
this, you must add the data values together to determine the data command value for each
port, which is the second number in the Test Set I/O or GPIB command. Refer to
Example 1
: (Option 129)
If the ports have
the same address
,
only one command is
needed.
P
ort 3 is the Source and Port
7
is the Receiver.
Source Port 3 = address 0, data 1, and Receiver Port
7
= address 0, data
48
.
The data values are added together, for an entry of 0.
49
.
For more information, refer to the PNA-X Help menu.
U3022AH10 User's and Service Guide
49
Содержание U3022AH10
Страница 7: ...U3022AH10 U3022AH10 Option 129 U3022AH10 User s and Service Guide 7...
Страница 9: ...U3022AH10 Introduction Figure 3 PNA X with U3022AH10 Option 129 U3022AH10 User s and Service Guide 9...
Страница 30: ...System Configurations U3022AH10 Figure 18 Test Set RF Path Diagram 30 U3022AH10 User s and Service Guide...
Страница 91: ...U3022AH10 User s and Service Guide 91...