U3022AH10
De
finitions
•
Specifications
describe the performance of parameters covered by the product warranty
(temperature –0 to 55 °C, unless otherwise noted.)
•
Typical
describes additional product performance information that is not covered by the
product warranty. It is performance beyond specification that 80% of the units exhibit with a
95% confidence level over the temperature range 20 to 30 °C. Typical performance does not
include measurement uncertainty.
•
Nominal
values indicate expected performance or describe product performance that is
useful in the application of the product, but is not covered by the product warranty.
•
Characteristic Performance
describes performance parameter that the product is expected to
meet before it leaves the factory, but is not verified in the field and is not covered by the
product warranty. A characteristic includes the same guard bands as a specification.
Available Options
Test Set Options
The U3022AH10 has two available options
:
Option 129
add
s an external source path for Ports 1, 3, 5, 7, 9, and 11. The
instrument cabinet is 498 mm deep and 221.5 mm high.
Option 700
is the standard model. The instrument cabinet is 421.6 mm deep and
177 mm high.
*
N
ew dark color version (Keysight Palette 2015)
The
must be installed into this network
analyzer file directory: c:\Program Files\Keysight\Network Analyzer\testsets
Definitions
Table
3
Interface Kit Options
12
/14
Port Network Analyzer System
Interface Kit
Option
Hardware Lock-Link
N5242A/B PNA-X
U3021-600021
U3022AH10 Opt 042
Cable Kit, Connector Type
U302
1
-600962, SMA m/m
Rack Mounting Kits
U3022AH10-1CM, Rack Mount Kit without Handles (1CM114A*) 5063-9216
U3022AH10-1CN, Front Handle Kit (1CN103A*) 5063-9229
U3022AH10-1CP, Rack Mount Kit with Handles (1CP106A*) 5063-9236
Network Analyzer Interface Kit Options
The
U3022AH10 requires the following kit to interface the test set with your
analyzer. The interface kit model option includes the hardware Lock-Link and
cable kit listed in
1.
"System Setup N5242A/B PNA-X" on page 2
2.
System Setup, 12/14 Port RF Cable Interface Configuration
12
U3022AH10 User's and Service Guide
Содержание U3022AH10
Страница 7: ...U3022AH10 U3022AH10 Option 129 U3022AH10 User s and Service Guide 7...
Страница 9: ...U3022AH10 Introduction Figure 3 PNA X with U3022AH10 Option 129 U3022AH10 User s and Service Guide 9...
Страница 30: ...System Configurations U3022AH10 Figure 18 Test Set RF Path Diagram 30 U3022AH10 User s and Service Guide...