Figure 6
6
A or B IN to Ports 3-12 Path Response, Standard
Table 2
1
U3022AE10
Troubleshooting the Test Set
Connect the RF Flex cables to the Test Port and A/B IN port indicated in Table 2
1
.
The expected results should b
e
similar to Figure 6
6
.
Source Signal Path Test Instructions
Path # RF Path
Description
Control Mode
<Address>.<Data>
Path
Components
Insertion Loss
(typical)
1
A IN to Port 3
0.16
P3 CPLR, S11
2
A IN to Port 5
0.32
P5 CPLR, S11
3
A IN to Port 7
0.48
P7 CPLR, S11
4
A IN to Port 9
0.64
P9 CPLR, S11
5
A IN to Port 11
0.80
P11 CPLR, S11
6
B IN to Port 4
16.16
P4 CPLR, S21
7
B IN to Port 6
16.32
P6 CPLR, S21
8
B IN to Port 8
16.48
P8 CPLR, S21
9
B IN to Port 10
16.64
P10 CPLR, S21
10
B IN to Port 12
16.80
P12 CPLR, S21