Hardware Lock-link Kit Installation (U3021-60002)
7. Place the PNA or PNA-X on top of the Test Set and ensure that the front frame of
the analyzer is positioned slightly forward of the locks that are attached to the
Test Set. Slide the analyzer back so the locks engage the front frame of the
analyzer.
Figure 1
5
Locking the Analyzer
8. Secure the PNA or PNA-X lower lock-links to the Test Set upper lock-link, using
the spring–loaded screws. If the analyzer's lock-link are not aligned with the
screw holes, loosen the screws securing the lock-link to the instrument slightly to
align and tighten.
Figure 1
6
Lock-Link Screws
The lock-links kit (U3021-60002) includes hardware to attach the PNA or PNA-X to
the test set. Refer to
“Network Analyzer Interface Kit Options” on page 13
and
“Contacting Keysight” on page 9
for ordering information.
• PNA or PNA-X- 5023-0132 (Kit includes lock-links and screws)
•
Test Set - N5242-20138 (right foot) and N5242-20139 (left foot) and screw
(0515-2317) for use with the PNA or PNA-X.
Test Set
Front Frame
Network Analyzer
Front Frame
Slide the network Analyzer
back to engage the lock link
in the back of the network
analyzer's front frame
Lock Link
screw
Slide the network analyzer
back to engage the lock link in
the back of the network
analyzer’s front frame,
26
Keysight U3022AE10 User's and Service Guide