Troubleshooting the Test Set
RF Switching Failures
If the test set’s RF signal functionality is not operating properly, use the following
procedure to aid in isolating the source of problem. The internal RF mechanical
switches used in this instrument are biased with the +21 Vdc supply voltage from the
Switch Interface board.
1.
Follow the
, if you have not done so.
2.
Place the network analyzer and test set side by side (not stacked) and connect
the I/O cable (N4011-21002). See
.
3.
Ensure that the network analyzer is
NOT
in Multiport Mode.
4.
Using the test set I/O Data command values, confirm the correct address and
data values are used, refer to
“Address and Data Values” on page
.
RF Switching Path Test
If you suspect an RF signal path problem with the test set and have verified that the
problem in not the front panel RF interface cables, the following procedure will check
all of the RF signal paths through the test set.
• Install the Test Set I/O cable from the analyzer to the test set’s rear panel.
•
Remove the front panel RF interface cables and reinstall the analyzer’s front panel
semi-rigid "U" shaped jumper cables.
Equipment Required
• PNA, PNA-L or PNA-X Network Analyzer
• Two RF Flex Cables (3.5 mm male/female)
• 3.5 mm Adapters (female to female)
•
Test Set I/O Cable
80
Keysight U3022AE10 User's and Service Guide