Step 6: Make a Measurement
Communications
Step 6: Make a Measurement
1. The first step in this process is to open the SFP of the M9381A and the M9393A.
a. Open the M9381A SFP by selecting Start > All Programs > Keysight >
M938x > M9381 SFP.
b. Open the M9393A SFP by selecting Start > All Programs > Keysight >
M9393 > M9393 SFP.
c. For each SFP, you are presented with the "Connect to Instrument" dia-
log. Use Ctrl/Select to select all of the modules that are components of
the M9381A and the M9393A and press Connect.
2. Connect a high quality SMA (male) to SMA (male) cable between the RF Out
connector on the Keysight M9310A PXIe Source Output and the RF In con-
nector on the M9365A.
3. Torque the connectors to 8 Lb-In (0.904 Nm).
4. On the M9381A SFP make the following settings:
a. Frequency: 2 GHz
b. Amplitude: -10 dBm
c. RF On: checked
38
Keysight Y1299A-004 RF PA/FEM Reference Solution Startup Guide
Содержание RF PA/FEM