General Power Meter Functions
2
Keysight N8262A User’s Guide
67
Figure 2-20
Bandwidth filter shapes
When video bandwidth is set to Off, it removes all digital signal conditioning. This
provides approximately 3 dB roll off at the maximum sensor’s bandwidth and is
best suited for accurate trace capture, minimizing overshoot, and removing any
ringing effects caused by the sharp cut-off filters used in the Low, Med and High
settings.
shows the filter shapes associated with the P-Series
Wideband Power Sensors.
Procedure
Set the video bandwidth as follows:
1
Click
. Click
to select the required channel.
2
Use the
and
softkeys to select the
Video B/W
setting field.
–3.5 dB
5 MHz 15 MHz
30 MHz
0
Frequency
Amplitude
P-Series Sensors
Approx
High
Med
Low
Off
Meter B/W
Setting
Channel