Maintenance
4
Keysight N8262A User’s Guide
137
–
330
Self-test Failed; ROM Check Failed
–
330
Self-test Failed; RAM Check Failed
–
350
Queue overflow
The error queue is full and another error has occurred which could not be recorded.
–
361
Parity error in program
The serial port receiver has detected a parity error and consequently, data integrity
cannot be guaranteed.
–
362
Framing error in program
The serial port receiver has detected a framing error and consequently, data integrity
cannot be guaranteed.
–
363
Input buffer overrun
The serial port receiver has been overrun and consequently, data has been lost.
–
410
Query INTERRUPTED
A command was received which sends data to the output buffer, but the output buffer
contained data from a previous command (the previous data is not overwritten). The
output buffer is cleared when power has been off, or after *RST (reset) command has
been executed.
–
420
Query UNTERMINATED
The power meter was addressed to talk (that is, to send data over the interface) but a
command has not been received which sends data to the output buffer. For example
you may have executed a CONFigure command (which does not generate data) and
then attempted to read data from the remote interface.
–430
Query DEADLOCKED
A command was received which generates too much data to fit in the output buffer and
the input buffer is also full. Command execution continues but data is lost.
–440
Query UNTERMINATED after indefinite response
The *IDN? command must be the last query command within a command string.
Содержание N8262A P Series
Страница 1: ...Keysight N8262A P Series Modular Power Meter User s Guide ...
Страница 18: ...18 Keysight N8262A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 20: ...20 Keysight N8262A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 106: ...2 General Power Meter Functions 106 Keysight N8262A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 146: ...5 Characteristics and Specifications 146 Keysight N8262A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...