General Power Meter Functions
2
Keysight N8262A User’s Guide
69
Setting Measurement Channel Gates
A system of gates, controlled by and referenced to a trigger point, is used to
obtain measurement data from a captured trace. The trace data within each gate
period is subsequently used for the individual measurement calculations. Up to 4
gates can be set up for each channel.
shows an example of 4 gates
setup to perform the following measurements simultaneously:
Average power level of the pulse:
Gate 1, average measurement
Average “off” power level ahead of the pulse:
Gate 2, average measurement
Peak to average ratio:
Gate 1, peak-to-average measurement
Pulse drop:
Gate 3, average measurement, minus Gate 4, average measurement
Figure 2-22
Example of measurement gates
NOTE
This feature is only available when a P-Series Power Sensor is connected.
Gate 3
Trigger
Gate 2
Gate 4
Gate 1