4
Maintenance
136
Keysight N8262A User’s Guide
–
310
System error; Dty Cyc may impair accuracy with ECP sensor
This indicates that the sensor connected is for use with CW signals only.
–
310
System error; Ch A Dty Cyc may impair accuracy with ECP sensor
This indicates that the sensor connected to Channel A is for use with CW signals only.
–
310
System error; Ch B Dty Cyc may impair accuracy with ECP sensor
This indicates that the sensor connected to Channel B is for use with CW signals only.
–
310
System error; Sensor EEPROM Read Failed - critical data not found or unreadable
This indicates a failure with your E-Series Power Sensor. Refer to your power sensor
manual for details on returning it for repair.
–
310
System error; Sensor EEPROM Read Completed OK but optional data block(s) not
found or unreadable
This indicates a failure with your E-Series Power Sensor. Refer to your power sensor
manual for details on returning it for repair.
–
310
System error; Sensor EEPROM Read Failed - unknown EEPROM table format
This indicates a failure with your E-Series Power Sensor. Refer to your power sensor
manual for details on returning it for repair.
–
310
System error; Sensor EEPROM < > data not found or unreadable
Where < > refers to the sensor data block covered, for example, Linearity, Temp - Comp
(temperature compensation).
This indicates a failure with your E-Series Power Sensor. Refer to your power sensor
manual for details on returning it for repair.
–321
Out of memory
The power meter required more memory than was available to run an internal
operation.
–
330
Self-test Failed;
The -330, “Self-test Failed” errors indicate that you have a problem with your power
meter. Refer to
“Contacting Keysight Technologies”
on page 139 for details of what to
do with your faulty power meter.
–
330
Self-test Failed; Measurement Channel Fault
–
330
Self-test Failed; Measurement Channel A Fault
–
330
Self-test Failed; Measurement Channel B Fault
–
330
Self-test Failed; Calibrator Fault
Refer to
on page 128 if you require a description of the calibrator test.
Содержание N8262A P Series
Страница 1: ...Keysight N8262A P Series Modular Power Meter User s Guide ...
Страница 18: ...18 Keysight N8262A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 20: ...20 Keysight N8262A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 106: ...2 General Power Meter Functions 106 Keysight N8262A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 146: ...5 Characteristics and Specifications 146 Keysight N8262A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...