Troubleshooting Guide
4
Keysight N1911A/1912A Service Guide
57
Communication Interface Failures
Type of communication
Debug tips
Possible faults
GPIB communication
Check/reseat the ribbon cable
connecting the PPMC to the main
board
Ribbon cable (low probability)
PPMC assembly (low probability)
Main board (high probability)
LAN/USB communication
Check visually to see whether or
not the connector is
obstructed/damaged
PPMC assembly