1 mW Power Reference Level Test
Description
The 1 mW power reference is used for the calibration of 8480 Series, N8480 Series
and E-Series power sensors, and is traceable to national standards. This test uses
an 8482A power sensor to transfer the power measured on an accurately
calibrated E4419B or E4417A power meter to the DUT reference.
Equipment
– Required test equipment:
1 unit of E4419B or E4417A dual channel power meter
1 unit of 8482A power sensor
11730A power sensor cable
– Either of these E4419B or E4417A power meters can be used. This specific
power sensor model must be used.
Figure 2-1
1 mW power reference level test setup connection diagram
Ch A
POWER METER
P-SERIES
POWER METER
NOTE
For rear panel options, the connections will differ from the illustration shown
here. Refer to the connector identification markings on the rear panel for further
details