Repair Guide
102
Keysight N1911A/1912A Service Guide
Step 7:
– Remove the 3 screws that attach the calibrator
assembly to the display support molding, and
separate them from one another
Note:
– Take care not to damage the EMI spring fingers
on the calibrator assembly
Step 8:
– Disconnect the backlight cable assembly from
the display interface board and inverter board
Step 9:
– Remove the 2 screws that attach the display
inter face board to the inverter board, and
separate them from one another
Instructions
Visual
EMI spring
finger