4
Troubleshooting Guide
54
Keysight N1911A/1912A Service Guide
Extended Self-Test
Instrument
Purpose
Debug tips
Possible faults
Keyboard
Verifies the operation of every
key (apart from the power
button)
Not applicable
Front panel (defective keymat
or key flex circuit)
Bitmap display
Verifies that all pixels in the
display can be illuminated in
various colors
Not applicable
Front panel (defective display,
display interface board, or
inverter board)
Time base
Provides a means to measure
time base frequency accuracy
Check that the BNC cable
being used is not damaged
Check that the BNC is
connected to ‘Trig Out’, not
‘Trig In’
Main board