Keysight M8132A 640 GSa/s Digital Signal Processor User’s Guide
65
Remote Programming
6
:SYSTem:VERSion?
Command
:SYSTem:VERSion?
Description
The query returns a formatted numeric value corresponding to the SCPI
version number for which the instrument complies.
Example
Query
:SYST:VERS?
:SYSTem:COMMunicate:*?
Command
:SYSTem:COMMunicate:*?
Description
The query returns information about the instrument Soft Front Panel’s
available connections. If a connection is not available, the returned value
is -1.
This is only useful if there is more than one Keysight module connected to
a PC, otherwise one would normally use the default connections (HiSLIP
and VXI-11 instrument number 0, socket port 5025, telnet port 5024)
One can never be sure if a socket port is already in use, so one could e.g.
specify a HiSLIP number on the command line (AgM8132SFP.exe /AutoID
/Inst5 /FallBack /r …) and let the Soft Front Panel find an unused socket
port. Then this socket port can be queried using the HiSLIP connection.
Example
Query
:SYST:COMM:*?
:SYSTem:COMMunicate:INSTr[:NUMBer]?
Command
:SYSTem:COMMunicate:INSTr?
Description
The query returns the VXI-11 instrument number used by the Soft Front
Panel.
Example
Query
:SYST:COMM:INST?
Содержание M8132A
Страница 1: ...User s Guide Keysight M8132A 640 GSa s Digital Signal Processor...
Страница 8: ......
Страница 16: ......
Страница 22: ......
Страница 46: ......
Страница 48: ......
Страница 94: ......
Страница 102: ......
Страница 106: ...This information is subject to change without notice Keysight Technologies 2020 Edition 2 0 May 2020 www keysight com...