Calibrating and Adjusting
3
2000/3000 X-Series Oscilloscopes Service Guide
59
Use 50
Ω
RG58AU or equivalent BNC cables when performing User Cal.
a
For a 2-channel oscilloscope, connect a BNC tee to the equal length cables.
Then connect a BNC(f)-to-BNC(f) (also called a barrel connector) to the tee
as shown below.
Longer cable
To Channel 1
To Channel 2
to TRIG OUT