Testing Performance
2
2000/3000 X-Series Oscilloscopes Service Guide
37
c
Press the
[Acquire]
key.
d
Then press the
Acq Mode
softkey and select
Averaging
.
e
Then press the
#Avgs
softkey and set it to 64.
Wait a few seconds for the measurement to settle.
4
Press the
[Cursors]
key, set the
Mode
softkey to
Normal
, then press the
X Y
softkey
and select
Y
. Press the
Y1
softkey, then use the Entry knob (labeled
on the
front panel) to set the Y1 cursor on the baseline of the signal.
5
Use the BNC tee and cables to connect the precision source /power supply to
both the oscilloscope and the multimeter (see
Table 12
Settings Used to Verify Dual Cursor Accuracy, 3000 X-Series Models
Volts/Div Setting
Precision Source Setting
Test Limits
5 V/Div
35 V
34.0 V
to
36.0 V
2 V/Div
14 V
13.6 V
to
14.4 V
1 V/Div
7 V
6.8 V
to
7.2 V
500 mV/Div
3.5 V
3.4 V
to
3.6 V
200 mV/Div
1.4 V
1.36 V
to
1.44 V
100 mV/Div
700 mV
680 mV
to
720 mV
50 mV/Div
350 mV
340 mV
to
360 mV
20 mV/Div
140 mV
136 mV
to
144 mV
10 mV/Div
70 mV
68 mV
to
72 mV
5 mV/Div
1
35 mV
34 mV
to
36 mV
2 mV/Div
1, 2
14 mV
13.2 mV
to
14.8 mV
1 mV/Div
1, 2
7 mV
6.2 mV
to
7.8 mV
1
A blocking capacitor is required at this range to reduce noise. See
2
Full scale is defined as 32 mV on the 2 mV/div range and the 1 mV/div range. Full scale on all other
ranges is defined as 8 divisions times the V/div setting.
Содержание InfiniiVision 2000
Страница 1: ...Keysight InfiniiVision 2000 3000 X Series Oscilloscopes Service Guide...
Страница 8: ...8 2000 3000 X Series Oscilloscopes Service Guide Contents...
Страница 10: ...10 2000 3000 X Series Oscilloscopes Service Guide Figures...
Страница 12: ...12 2000 3000 X Series Oscilloscopes Service Guide Tables...
Страница 62: ...62 2000 3000 X Series Oscilloscopes Service Guide 3 Calibrating and Adjusting...
Страница 72: ...72 2000 3000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies...