22
Keysight UXR-Series and MXR/EXR-Series MultiScope Hardware Configuration Guide
3
Three- to Five-Oscilloscope Systems, UXR-Series
Before connecting the follower #2 through follower #4 oscilloscopes, first follow
the instructions for connecting a two-oscilloscope rack mount system in
“Connecting the Oscilloscopes"
Then, for each of the oscilloscopes from follower #2 to follower #4, using the parts
in a N2165A (add one) expansion kit, follow these instructions:
1
Remove the SMA 50
Ω
terminator (1250-3745) from the power divider output
that will be used for the oscilloscope.
2
Connect one 54964-61632 cable between the power divider output and the
oscilloscope’s
Sample_CLK_In
input.
3
Connect one 54964-61632 cable between the previous oscilloscope’s
Trig Out
connector to this oscilloscope’s
Aux Trig In
connector.
4
Connect one 54964-61632 cable between the previous oscilloscope’s
Reference CLK Out
connector to this oscilloscope’s
Reference CLK In
connector.
5
Connect the oscilloscope to the Leader oscilloscope (or external host
computer) through the desired remote interface connection (LAN, USB, etc.).
6
Содержание EXR Series