5
Bring up the M8195A SFP. Select the
Clock
tab and enter the
Reference Clock
Frequency
as
10 MHz
.
Change the position of the two switches as shown in the
. This step
will ensure that the signal generation and measurement system’s timing base
are synchronized from the N5183Bs’ 10 MHz reference output.
Figure 21
Switch 10 MHz path using M8195A SFP
6
Set the RF frequency and amplitude of the
N5183B
. Press the
FREQ
button
and enter
37.5 GHz
. Press the
AMPTD
button, enter
6 to 8 dBm
(depends on
LO cable loss). Turn on the source output.