Keysight E8740A-060 Start-up Guide
25
3
Set center frequency and span to
5 GHz.
a
Select the
Frequency
panel or press
the
FREQ
hard key.
b
In the
Center Frequency
control, enter
79.0 GHz.
c
In the
Span
control enter the
maximum span, 1 GHz.
The maximum span allowable depends
on the N9041B bandwidth option.
4
Set full range and mechanical
attenuation to 0 dB.
a
Select the
Amplitude
panel or press
the
AMPTD
key.
b
Select the
Attenuation
tab and tap the
Full Range Atten
control. Enter 0 dB
using the soft numeric keypad.
c
Tap the
Mech Atten
control and enter
0 dB with the soft numeric keypad.
Or set the attenuation level to ensure the
input signal does not exceed mixer limit.
5
Switch the output to the rear external
wideband IF output connector.
a
Select the
Input/Output
panel.
b
Choose the
Output
Tab and toggle the
Ext IF Out
switch to
On
.
No signal will appear on the SA screen
because the signal has been diverted to
the rear connector.
A warning message will appear on the
Spectrum Analyzer as shown in
Step
Action
Notes
Содержание E8740A-060
Страница 4: ...4 Keysight E8740A 060 Start up Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 6: ...6 Keysight E8740A 060 Start up Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 11: ...Keysight E8740A 060 Start up Guide 11 Figure 1 Average noise density marker noise at 79 GHz Step Action Notes...
Страница 23: ...Keysight E8740A 060 Start up Guide 23 Figure 10 Out of band spurious sweep Step Action Notes...
Страница 34: ...34 Keysight E8740A 060 Start up Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...