10
Keysight E8740A-060 Start-up Guide
7
Set the detector average type to
RMS
.
a
Click on the top horizontal bar of the
N9041B interface.
b
Select
Average Type
and choose
Power (RMS)
from the drop-down list.
8
Measure the average noise level from
74 to 84 GHz.
a
Select the
Marker
panel or press the
Marker
key.
b
Select the
Settings
tab and in the
Marker Frequency
control, enter
79.0 GHz.
c
Select
Marker Function
tab and then
the
Band Function
control. Choose
Marker Noise
. Set
Band Span
to
10 GHz.
d
To turn marker off set
Marker Mode
to
Off
.
below displays the Average
Noise Density Marker at 79 GHz band
with a value close to -150 dBm/Hz.
Step
Action
Notes