Characteristics - Measured: DC and AC Current Burden Voltage at Full Scale.
DC current range
Burden voltage
1 µA
< 0.0011 V
10 µA
< 0.011 V
100 µA
< 0.11 V
1 mA
< 0.11 V
10 mA
< 0.027 V
100 mA
< 0.27 V
1 A
< 0.7 V
Characteristics - Nominal: Other
A/D linearity: 0.0001% of r 0.0001% of range (measured)