DAQM904A 

4x8 Two-wire Matrix Switch

 

– 32 two-wire crosspoints

 

– 300 V, 1 A switching

The Keysight DAQM904A gives you the most flexible connection 
path between your device under test and your test equipment, 
allowing different instruments to be connected to multiple points 
on your DUT at the same time. 

Rows or columns may be connected between multiple modules to 
build 8x8, 4x16 or larger matrices, with up to 96 crosspoints in a 
single frame.

Row 1

Row 2

Row 3

Row 4

Channel 31

(Row 3, Column 1)

Col 1

 

H     L

Col 2

 

H     L

Col 8

 

H     L

H
L

H
L

H
L

H
L

H

H

L

L

DAQM905A 

Dual 4-Channel RF Multiplexer 50 

 

– 2 GHz bandwidth 

 

– BNC to SMB adapter cables included

The Keysight DAQM905A RF multiplexers offer broadband 
switching capabilities for high frequency and pulsed signals. Use 
them to route test signals between your device under test and 
your signal generator, oscillo-scope, spectrum analyzer, or other 
instrumentation.

The RF multiplexer are arranged as two independent 1x4 
multiplexers, each with a common shield and a switched center 
conductor. Connections can be made directly to SMB inputs 
with 2 GHz usable bandwidth, or to the BNC-to-SMB adapters 
provided with 1 GHz bandwidth. Multiple banks may be cascaded 
together for applications requiring even larger topologies—create 
a 16:1 multiplexer in a single frame. 

11

12

Com

13

14

21

22

Com

23

24

www.allice.de 

Allice Messtechnik GmbH 

Содержание DAQ970A

Страница 1: ...T E C HNI C A L O V E R V IE W Keysight DAQ970A Data Acquisition System www allice de Allice Messtechnik GmbH...

Страница 2: ...tioning and modular flexibility all in a low cost compact data acquisition package The DAQ970A features 6 digits 22 bits of resolution 0 004 basic dcV accuracy and ultra low reading noise Combine that...

Страница 3: ...tection and LAN lock out are also provided to limit access The graphical Web interface makes it easy to configure measurements set up and execute scans or trouble shoot your designs from anywhere on t...

Страница 4: ...tiation of the measurements 6 Measurement configuration menu to set parameters for measurements 7 Knob 8 Cursor navigation keypad Label Description 1 Slot identifier 100 200 300 2 External trigger inp...

Страница 5: ...A is easy to use for a multitude of data logging and monitoring applications either stand alone or with a computer Its flexible modular design makes it scalable from 20 to 120 channels and lets you ad...

Страница 6: ...r later uploading to a PC Or if you need even more memory the DAQ970A USB port can be used to log data directly to a USB flash drive or to copy the data from the reading memory without being connected...

Страница 7: ...are drivers that support C C Visual Basic Visual Studio LAN and USB on the DAQ970A and SCPI programming language make integration even easier The functionality you need We put a lot of thought into de...

Страница 8: ...rement switching and control capabilities to choose from Modules include both low frequency and RF multiplexers a matrix switch a general purpose switch and a multifunction module that includes digita...

Страница 9: ...ing errors are also listed separately in the module specifications section Temperature measurement accuracies include ITS 90 conversion errors The thermocouple accuracies include the reference junctio...

Страница 10: ...0 0006 0 0040 0 0006 0 0055 0 0006 0 0005 0 0001 0 0002 0 0001 300 V 0 0020 0 0020 0 0035 0 0020 0 0040 0 0020 0 0055 0 0020 0 0005 0 0001 0 0002 0 0001 Resistance2 7 100 0 0030 0 0050 0 0050 0 0060 0...

Страница 11: ...Hz 10 Hz 0 10 0 05 0 10 0 06 0 10 0 06 0 11 0 06 0 010 0 008 10 Hz 20 kHz 0 02 0 05 0 04 0 06 0 05 0 06 0 06 0 06 0 010 0 008 20 kHz 50 kHz 0 05 0 09 0 06 0 09 0 07 0 09 0 08 0 09 0 010 0 0012 50 kHz...

Страница 12: ...uple DAQ970A extended temp ranges Probe5 Ref junction DMM accuracy E 200 C to 150 C 1 4 C J 210 C to 150 C 1 6 C K 200 C to 100 C 1 7 C N 200 C to 100 C 2 C R 50 C to 300 C 2 9 C T 200 C to 100 C 1 7...

Страница 13: ...20 pF 50 pF Capacitance Lo Earth NA 150 pF 135 pF 80 pF 80 pF NA 80 pF Volt Hertz limit 10 8 10 8 10 8 10 8 10 8 10 10 10 8 Characteristics typical Other T C CRJ accuracy Switch life no load Unlimited...

Страница 14: ...ms max alarm sampling Latency 5 ms typical to DAQ970A alarm output Read write speed 95 s Max count 2 26 1 Totalize input 100 kHz max Rising or falling edge programmable min H or L time 5uS Signal leve...

Страница 15: ...N 275 275 907A Totalizer 240 240 Characteristics Nominal Data out of Memory FETCh of 50K readings Single channel DAQ970A over USB DAQ970A over LAN Readings 55000 s 120000 s Rdgs with timestamp 35000 s...

Страница 16: ...t Burden Voltage at Full Scale DC current range Burden voltage 1 A 0 0011 V 10 A 0 011 V 100 A 0 11 V 1 mA 0 11 V 10 mA 0 027 V 100 mA 0 27 V 1 A 0 7 V Characteristics Nominal Other A D linearity 0 00...

Страница 17: ...ion EN 61010 2 030 2010 1st Edition ANSI ISA 61010 2 030 82 02 03 First Edition ANSI UL 61010 2 030 First Edition CAN CSA C22 2 No 61010 2 030 First Edition Refer to Declaration of Conformity for curr...

Страница 18: ...rature Operating all cards 0 to 55 C all cards 40 to 70 C all cards 40 C to 80 RH Storage Humidity non condensing 1 Not recommended for connection to ac line without external transient suppression Up...

Страница 19: ...n 120 V switching The DAQM900A module is divided into two banks of 10 two wire channels each All 20 channels switch both HI and LO inputs thus providing fully isolated inputs to the internal DMM or an...

Страница 20: ...ls 22 Com Current Fuse Fuse Internal DMM input 4W sense Internal DMM input Current Reference junction sensor Channel switches Bank switch Shunt switches DAQM902A 16 Channel High Speed Multiplexer Scan...

Страница 21: ...rnal measurements except current are supported The module low connection is isolated from earth and can float up to 300 V DAQM903A 20 Channel Actuator General Purpose Switch SPDT Form C latching relay...

Страница 22: ...el RF Multiplexer 50 2 GHz bandwidth BNC to SMB adapter cables included The Keysight DAQM905A RF multiplexers offer broadband switching capabilities for high frequency and pulsed signals Use them to r...

Страница 23: ...allows great flexibility for a variety of sense and control applications It combines two 8 bit ports of digital input and output a 100 kHz gated totalizer and two 12 V or 24 mA analog outputs all on a...

Страница 24: ...e With 26 bits of resolution it can count events at full speed for nearly 11 minutes without an overflow Max count 226 1 Totalize input 100 kHz max Rising or falling edge programmable Signal level 1 V...

Страница 25: ...mensions To rack mount two instruments side by side order thelock link kit part number DAQA194A FG and flange kit part number DAQA191A FG To rack mount a single instrument order the adapter kit part n...

Страница 26: ...nnel Solid state multiplexer DAQM901A 20 Channel armature multiplexer DAQM902A 16 Channel reed multiplexer DAQM903A 20 Channel actuator general purpose switch DAQM904A 4 x 8 Two wire matrix switch DAQ...

Отзывы: