N9069C Noise Figure Measurement Guide 67
Measurement Related Tasks and Concepts
Setting the Signal Analyzer to Make a Measurement
4.
Press
Edit Meas Table
to access the ENR table for
which you wish to enter data.
An ENR Table appears with a table editing and
navigation menu on the left side. See the following
figure.
To clear an existing table, press
Clear Table
.
To enter common measurement and calibration
ENR data, press
Use Meas Table Data for
Cal
to
On
. The measurement table data is
automatically used for the calibration data.
To enter either measurement ENR data or
calibration ENR data, make sure that
Use Meas
Table Data for Cal
is set to
Off
, and then
select your table by pressing either
Meas Table
or
Cal Table
.
Figure 3-5
ENR Table
5.
(Optional) Press
Serial #
, enter the noise source
serial number using the on screen keyboard or an
external USB keyboard, and press
Done
.
6.
(Optional) Press
Model ID
, enter the noise source
model number using the on screen keyboard or an
external USB keyboard, and press
Done
.
7.
Press
Go To Row
to move the highlight to the
Frequency/ENR column. Enter the values in the table
using the numeric keys and terminate it using the
unit menu keys.
You can insert the frequencies into the ENR
Table entry in any order, as the analyzer
automatically sorts the frequency list into
ascending order.
Table 3-2
Step
Notes