N9069C Noise Figure Measurement Guide 73
Measurement Related Tasks and Concepts
Setting the Signal Analyzer to Make a Measurement
Using Fixed Frequency Mode
The fixed frequency mode is used when you want to make a measurement at a
single frequency.
To set a fixed frequency, press
FREQ
,
Freq Mode
,
Fixed
, then press
Fixed Freq
and enter the frequency value using the numeric keys and the unit termination
menu.
Setting the Bandwidth and Averaging
Effect of Bandwidth and Averaging on Speed, Jitter, and Measurement
Accuracy
Jitter is a natural occurrence when measuring noise. To reduce jitter you must
increase the number of averages or increase the measurement bandwidth.
If the bandwidth is reduced, you need to increase the number of averages to
maintain the same uncertainty.
The greater the number of averages chosen, the more accurate the
measurement, as this reduces jitter on the measurement. However, this must
be considered against how long it takes to complete the measurement.
There is therefore a trade off between speed and the accuracy and uncertainty
of a measurement.
Selecting the Resolution Bandwidth (RBW) Value
To set the RBW value, press
BW
,
RBW
and select whether the resolution
bandwidth is to be set automatically, or to be set manually by you.
When the resolution bandwidth is set to Auto (default setting), the bandwidth
is set automatically, and is dependent on the measurement frequency.
At measurement frequencies of 8 MHz or above, the resolution bandwidth is
set automatically to 4 MHz.
At measurement frequencies less than 8 MHz, the resolution bandwidth is set
automatically to approximately one third of the measurement frequency.
If you have not entered the noise source ENR data that you intend using for
the fixed frequency mode measurement, you may specify a spot ENR value
and set the ENR mode to Spot.