N9069C Noise Figure Measurement Guide 17
Making Noise Figure Measurements
Making Basic Measurements
10.
Set the BW
— Press
BW
, toggle the
Res BW
to
Auto
.
11.
Set the attenuation
— Press
MEAS SETUP
,
select the
Cal Setup
tab,
change the
Min Atten
and
Max Atten
values if
required.
This example uses the default minimum and
maximum input attenuation.
See
“Selecting the Input Attenuation
for more details.
12.
Perform a calibration
— Press
CALIBRATE
.
You will be asked to confirm the calibration.
See the following figure for results after
calibration.
With the calibration completed and no device under test inserted, both gain and noise figure are near 0 dB. This shows that
the analyzer has removed the noise contribution from the measurement system.
13.
View the results in a table
— Press
Display
and select
Table
from the
Layout
drop down list.
A result similar to the following graphic is
now displayed. The expectation is
approximately 0 dB of noise figure and gain.
It may be better to view these results using
the table layout mode
Step
Action
Notes