5-18 Electronic Calibration Modules Reference Guide
Specifications and Characteristics
Characteristic Performance
5-
N469xD Series
Table 5-32 Characteristic Performance for N4690D (Type-N 50
Ω
)
Parameter
Frequency Range
DC to
2 MHz
a
a. For option 0DC.
2 MHz to
10 MHz
300 kHz
to 2 MHz
b
b. For option 003.
2 MHz
to 10 MHz
10 MHz to
500 MHz
500 MHz
to 2 GHz
2 GHz
to 10 GHz
10 GHz
to 18 GHz
Transmission
Tracking (±dB)
0.17
c
c. Values based on using the network analyzer N5231B Option 200.
0.06
0.37
0.08
0.10
d
d. Values based on using the network analyzer N5222B Option 200.
0.04
0.09
Load Match (dB) 36
26
33
39
Table 5-33 Characteristic Performance for N4691D (3.5 mm)
Parameter
Frequency Range
DC to
2 MHz
a
a. For Option 0DC.
2 MHz to
10 MHz
c
300 kHz
to
2 MHz
b
b. For Option 003.
2 MHz
to 10 MHz
10 MHz
to
500 MHz
500 MHz
to 2 GHz
2 GHz
to 10 GHz
10 GHz
to
20 GHz
20 GHz
to 26.5 GHz
Transmission
Tracking (±dB)
0.21
c
c. Values based on using the network analyzer N5231B Option 200.
0.06
0.37
0.08
0.09
d
d. Values based on using the network analyzer N5222B Option 200.
0.04
Load Match (dB) 34
46
40
38
Содержание 85091D
Страница 4: ...4 ...
Страница 44: ...1 36 Electronic Calibration Modules Reference Guide General Information Safety and Regulatory Information 1 ...
Страница 66: ...2 22 Electronic Calibration Modules Reference Guide Preparing ECal for Use Operating and Safety Precautions 2 ...
Страница 82: ...3 16 Electronic Calibration Modules Reference Guide Operating ECal Module ECal Troubleshooting 3 ...
Страница 170: ...6 14 Electronic Calibration Modules Reference Guide Replaceable Parts Other ECal Accessories 6 ...
Страница 171: ......