Keysight 85059B Operation and Service Guide 3-13
Use, Maintenance, and Care of the Devices
Making Connections
Figure 3-7
Alignment
Connector Misalignment
Forced
misalignment could damage the female center conductor.
Figure 3-8
Misalignment
How to Separate a Connection
To avoid lateral (bending) force on the connector mating plane surfaces,
always support the devices and connections.
1.
Use an open-end wrench to prevent the device body from turning.
2.
Use another open-end wrench to loosen the connector nut.
Turn the connector nut,
not
the device body. Major damage to the center
conductor can occur if the device body is twisted.