Keysight 85059B Operation and Service Guide 1-7
General Information
Offset-Short Calibration Using the 85059B (Precision) Calibration Kit and the PNA’s
Guided Power Calibration
Gage sets (not provided with the 85059B), adapters, ESD protection devices,
and va
rious
connector cleaning supplies are not included in the calibration kit
but are required to ensure successful operation of the calibration kit. Refer to
for ordering information.
Offset-Short Calibration Using the 85059B (Precision) Calibration
Kit and the PNA’s Guided Power Calibration
IMPORTANT!
The 85059B calibration kit must be used with the PNA’s
Guided Calibration. PNA’s guided calibration selects the proper standards
for connection based on the selected frequency range. Refer to
.
To download a copy of the 85059B calibration kit definitions (.xkt file), refer
to Keysight’s Calibration Kit Definitions Web page at
http://na.support.keysight.com/pna/caldefs/stddefs.html
For best measurement results, the isolation standard should be the
equivalent impedance of the DUT.