2-4 Keysight 85059B Operation and Service Guide
Specifications
Electrical Specifications
2-
Electrical Specifications
The electrical specifications in
apply to the devices in your
calibration kit when connected with a Keysight precision interface.
Table 2-2
85059B (Precision) Electrical Specifications for 1.0 mm
Devices
Device
Frequency
Parameter
(GHz)
Data-Based Model
Deviation from
Nominal Phase
(degrees)
Short 1, 2, 3, & 4
Male and Female
DC to 20
1.50
20 – 50
2.30
50 — 80
3.25
80 — 100
3.50
100 — 120
4.00
Open Male and
Female
DC to 20
2
20 – 50
3
50 — 80
7
80 — 100
8
100 — 120
8
Device
Frequency
Parameter
(GHz)
Return Loss (dB)
Load Male and
Female
DC to 18
30
18 to 40
26
40 to 50
24