3-10 Keysight 85055A User’s and Service Guide
Use, Maintenance, and Care of the Devices
Gaging Connectors
3-
d.
Read the gage indicator dial. If the needle has moved clockwise, the
center conductor is
protruding
by an amount indicated by the black
numbers. If the needle has moved counterclockwise, the center
conductor is
recessed
by an amount indicated by the red numbers.
For maximum accuracy, measure the connector a minimum of three
times and take an average of the readings. After each measurement,
rotate the gage a quarter-turn to reduce measurement variations
that result from the gage or the connector face not being exactly
perpendicular to the center axis.
e.
Compare the average reading with the specifications listed in
f.
Without turning either the gage or the device, remove the device
from the gage.