20
Place the 400
µ
m standard foil on the substrate and repeat
the measurement procedure 5 times.
• Depending upon the type of material, the value displayed may differ greatly
from that indicated in the example to the left. However, you can continue
the procedure by setting it near the setting value by performing steps
-
and
=
.
Press the
key while holding the probe toward the air.
"---" is shown in the display.
Input the coating thickness of the calibraiton foil (100.0
µ
m).
Press the
key. The number displayed disappears and S.
NO increments from 2 to 3 to indicate that the unit is ready
for calibration using the third standard foil.
^
&
*
<Calibration using the standard foil (400
µ
m)>
(
Step
Operation
Display
Explanation
(Measure 5 times)
1
2
5
○○○
○