16
Press the
ENT key
10 times.
Step
Operation
Display
Explanation
Press the
key.
Press the
key.
Press the
key 10 times while pressing the probe against
the substrate.
• The substrate should be an uncoated (no plating, paint, etc.) sample of
the same shape and material as that to be measured.
After pressing the
key 10 times the buzzer will sound
intermittently and "IF" will appear in the display.
Press the
key while holding the probe toward the air.
• The probe must be in the air, don’t touch any surface or others.
1
2
3
4
○○○○○○
○○○○
○
2) Substrate Calibration Procedure
Once substrate calibration settings have been registered it is not necessary to perform this operation each time a
measurement is made. However this procedure must be performed again when changes such as replacement of the
probe are made.
•
Example: Using an electromagnetic probe