17
⑥
Reset
Press the [
Tare/Reset
] key.
The display of the results of the measurement (i.e.,
moisture content) will disappear and be replaced by a
display of the weight of the sample after drying.
* Reset cannot be performed while a buzzer sounds.
⑦
Disposing of samples after measurement
Open the heater lid, remove the sample dish with the
sample dish handler, and dispose of measured sample.
* Note that the sample and sample dish may be very hot and
you should be careful in handling them at this time.
⑧
Preparing for the Next Measurement
Leave the lid to the heater open to allow the tester to cool.
When performing sequential measurements, make sure
that the entire unit has cooled before proceeding to the
next measurement.
You should also keep a spare (cooled) sample dish
available to use in the next measurement.
(See “■ Tips on how to ensure accurate measurements”
on page 18 for further information.)
* The guideline is for “the entire unit” is when 40ºC or lower
is displayed.
When ready, return to step
②
and begin the next
measurement.
⑨
Turning off the power
When all the measurements have been completed,
remove the power plug from the power outlet.
When not using the main unit, always remove the power
plug from the power outlet.