Maintenance
59 / 61
2019/11
010-917_SON_EN
Permanent odour development
Fault
Possible cause
Action
Odour pollution
Wastewater pipes leaking.
Check firm fit and seals, repair if necessary
No venting pipe, cross-section too
small
Retrofit on-site
Closed room with no air exchange
Create ventilation possibilities, forced ventilation
System parts are leaking
Eliminate leaks
8.5
Clean the grease separator
• Make sure that no more wastewater can be fed into it.
• Empty the system tank as described under “Emptying” (Chap5 on page 38).
• Disconnect the power supply.
•
Take both inspection covers off the system tank.
Do not clean the grease separator system using a water pressure of more than 5 bar and a water
temperature of more than 50°C. Do not use a high-pressure cleaner on seals. If soap is used for
cleaning, rinse out / extract the residue, as otherwise it could lead to functional problems.
• Clean all components with hot water.
• If present, clean the
SonicControl
sensor.
• Fit both inspection covers on the system tank.
•
Carry out the pressure test and subsequent functional check (see 3.4 Initial filling and pressure test on page
32)).
If all the system components are airtight the grease separator can be put into operation again.
Functions with
EasyClean
“Auto Mix & Pump“
Cleaning program:
You can adapt the separator cleaning (flushing with hot water) to the general local conditions using menu
Cleaning program
(see 1.6.14 on page 50).
Legionella flushing:
The drinking water pipes can be flushed automatically using the menu
Legionella flushing interval
(see 1.6.15 on
page 50).