16 Error messages
Er1 Hi
Initial weight error
Er2 nuLL
Value below allowed range
Er3 FuL1
Value above allowed range
Er4 FuL2
Weighing range exceeded
Er5 rout
Value outside allowed range
e.g. tare value <= 0,
Reference weight = 0
Er7 tout
Zero/Tare not possible as stable weighing value is still pending
Er8 outr
Input outside range
e.g. for tolerance control: Input upper limit<lower limit
Er9 Lock
Function blocked
Er10 cal
Adjustment error e.g. incorrect adjustment weight}
17 Service, maintenance, disposal
17.1 Cleaning
Before cleaning, please disconnect the appliance from the operating voltage.
Please do not use aggressive cleaning agents (solvents or similar agents), but a
cloth dampened with mild soap suds. Ensure that no liquid penetrates into the device
and wipe with a dry soft cloth.
Loose residue sample/powder can be removed carefully with a brush or manual
vacuum cleaner.
Spilled weighing goods must be removed immediately.
17.2 Service, maintenance
The appliance may only be opened by trained service technicians who are
authorized by KERN.
Before opening, disconnect from power supply.
17.3 Disposal
Disposal of packaging and appliance must be carried out by operator according to
valid national or regional law of the location where the appliance is used.